BAXTER SYRINGE PUMP SYRINGE ALARM PLACE A SYRINGE WITH THE PLUNGER AT
BAXTER SYRINGE PUMP MODEL INFUS OR,2L3100,6461500 REQUIRED TEST EQUIPMENT: SILICON LUBRICANT, PROPOFOL LABEL, 60CC SYRINGE, GRADUATED CYLANDER AND TIMER OR INFUSION PUMP TESTER LUBE THREADED DRIVE ROD – THE DRIVE ROD IS ACCESSABLE THRU THE SIDE CASE CHANNEL (BEHIND CLEAR RUBBER SEAL). USE SILICON LUBRICANT SUCH AS LEVEL ONE FLUID WARMER SILICON P/N # 80-04-002 OR GE VERSILUBE G322L GREASE. CHECK GENERAL CONDITION OF EQUIPMENT, SPECIFICALLY THE PUSHER BLOCK AND SYRINGE HOLDER ASSEMBLIES. END OF SYRINGE ALARM – PLACE A SYRINGE WITH THE PLUNGER AT 3ML INTO THE PUMP. START INFUSING AND VERIFY THAT WHEN THE SYRINGE IS AT APPROX. 1 ML THAT THERE IS AN AUDIBLE ALARM AND THAT THE PUMP STOPS INFUSING. OCCLUSION ALARM – WHILE THE PUMP IS INFUSING MOVE THE SYRINGE HOLDER DOWNWARD AND VERIFY THAT THERE IS AN AUDIBLE ALARM AND THAT THE PUMP STOPS INFUSING. CHECK RATE ACCURACY – FILL A 60 CC SYRINGE AND PRIME ALL AIR FROM IT. USING A PROPOFOL LABEL SET THE INFUSION RATE TO 300 MCG/KG/MIN AND BODY WIEGHT TO 100 KG. USING AN INFUSION PUMP TESTER OR A TIMER AND A GRADUATED CYLANDER VERIFY A RATE OF 3.0 ML/MIN +/-3%. CHECK VOLUME ACCURACY – USING A PROPOFOL LABEL SET THE INFUSION RATE TO 50 MCG/KG/MIN, BODY WIEGHT TO 100 KG AND BOLUS TO 1000 MCG/KG. INITIATE THE BOLUS AND VERIFY THAT THE VOLUME MEASURED IS 10 ML +/- 3 %. CHECK BOLUS DURATION – SET BOLUS TO 1000 MCG/KG, INITIATE A BOLUS DOSE AND VERIFY THE DELIVERY DURATION IS 1 MINUTE +/- 6 %. |