BAXTER SYRINGE PUMP SYRINGE ALARM PLACE A SYRINGE WITH THE PLUNGER AT




BAXTER SYRINGE PUMP

MODEL INFUS OR,2L3100,6461500 
 
 
 

REQUIRED TEST EQUIPMENT:  SILICON LUBRICANT, PROPOFOL LABEL,

                                                              60CC SYRINGE, GRADUATED CYLANDER

                                                              AND TIMER OR INFUSION PUMP TESTER 
 
 

LUBE THREADED DRIVE ROD – THE DRIVE ROD IS ACCESSABLE THRU

                        THE SIDE CASE CHANNEL (BEHIND CLEAR

                        RUBBER SEAL).  USE SILICON LUBRICANT SUCH

                        AS LEVEL ONE FLUID WARMER SILICON P/N #

                        80-04-002 OR GE VERSILUBE G322L GREASE.

CHECK GENERAL CONDITION OF EQUIPMENT, SPECIFICALLY THE PUSHER BLOCK AND SYRINGE HOLDER ASSEMBLIES.

 
 

END OF SYRINGE ALARM – PLACE A SYRINGE WITH THE PLUNGER AT

                        3ML INTO THE PUMP.  START INFUSING AND

                        VERIFY THAT WHEN THE SYRINGE IS AT APPROX.

                        1 ML THAT THERE IS AN AUDIBLE ALARM AND

                        THAT THE PUMP STOPS INFUSING. 
 

OCCLUSION ALARM – WHILE THE PUMP IS INFUSING MOVE THE SYRINGE

                           HOLDER DOWNWARD AND VERIFY THAT THERE IS

                                             AN AUDIBLE ALARM AND THAT THE PUMP STOPS

                                             INFUSING. 
 

CHECK RATE ACCURACY – FILL A 60 CC SYRINGE AND PRIME ALL AIR

                                               FROM IT.  USING A PROPOFOL LABEL SET THE

                                               INFUSION RATE TO 300 MCG/KG/MIN AND BODY

                                               WIEGHT TO 100 KG.  USING AN INFUSION PUMP

                                               TESTER OR A TIMER AND A GRADUATED

                                               CYLANDER VERIFY A RATE OF 3.0 ML/MIN +/-3%. 
 

CHECK VOLUME ACCURACY – USING A PROPOFOL LABEL SET THE

                        INFUSION RATE TO 50 MCG/KG/MIN, BODY

                        WIEGHT TO 100 KG AND BOLUS TO 1000 MCG/KG.

                        INITIATE THE BOLUS AND VERIFY THAT THE

                        VOLUME MEASURED IS 10 ML +/- 3 %. 
 

CHECK BOLUS DURATION – SET BOLUS TO 1000 MCG/KG, INITIATE A

                        BOLUS DOSE AND VERIFY THE DELIVERY

                        DURATION IS 1 MINUTE +/- 6 %.






Download links for : << Also incoming a flow premixed ethanol water flow controlled a Teledyne ISCO Syringe pump Outgoing >>
"BAXTER SYRINGE PUMP SYRINGE ALARM PLACE A SYRINGE WITH THE PLUNGER AT"


How to Download
You may need eMule or Bittorrent to download ebook torrents or emule links.

Report Dead Link
Please leave a comment to report dead links, so that someone else may update new links.


Search More...

[share-ebook]BAXTER SYRINGE PUMP SYRINGE ALARM PLACE A SYRINGE WITH THE PLUNGER AT

Google

Related Books


Books related to :

<< Also incoming a flow premixed ethanol water flow controlled a Teledyne ISCO Syringe pump Outgoing

capillary design because the process can already automated syringe pumps In addition it seemed >>


The New York Times rss

    Google

    BAXTER SYRINGE PUMP SYRINGE ALARM PLACE A SYRINGE WITH THE PLUNGER AT

    BAXTER SYRINGE PUMP

    MODEL INFUS OR,2L3100,6461500 
     
     
     

    REQUIRED TEST EQUIPMENT:  SILICON LUBRICANT, PROPOFOL LABEL,

                                                                  60CC SYRINGE, GRADUATED CYLANDER

                                                                  AND TIMER OR INFUSION PUMP TESTER 
     
     

    LUBE THREADED DRIVE ROD – THE DRIVE ROD IS ACCESSABLE THRU

                            THE SIDE CASE CHANNEL (BEHIND CLEAR

                            RUBBER SEAL).  USE SILICON LUBRICANT SUCH

                            AS LEVEL ONE FLUID WARMER SILICON P/N #

                            80-04-002 OR GE VERSILUBE G322L GREASE.

    CHECK GENERAL CONDITION OF EQUIPMENT, SPECIFICALLY THE PUSHER BLOCK AND SYRINGE HOLDER ASSEMBLIES.

     
     

    END OF SYRINGE ALARM – PLACE A SYRINGE WITH THE PLUNGER AT

                            3ML INTO THE PUMP.  START INFUSING AND

                            VERIFY THAT WHEN THE SYRINGE IS AT APPROX.

                            1 ML THAT THERE IS AN AUDIBLE ALARM AND

                            THAT THE PUMP STOPS INFUSING. 
     

    OCCLUSION ALARM – WHILE THE PUMP IS INFUSING MOVE THE SYRINGE

                               HOLDER DOWNWARD AND VERIFY THAT THERE IS

                                                 AN AUDIBLE ALARM AND THAT THE PUMP STOPS

                                                 INFUSING. 
     

    CHECK RATE ACCURACY – FILL A 60 CC SYRINGE AND PRIME ALL AIR

                                                   FROM IT.  USING A PROPOFOL LABEL SET THE

                                                   INFUSION RATE TO 300 MCG/KG/MIN AND BODY

                                                   WIEGHT TO 100 KG.  USING AN INFUSION PUMP

                                                   TESTER OR A TIMER AND A GRADUATED

                                                   CYLANDER VERIFY A RATE OF 3.0 ML/MIN +/-3%. 
     

    CHECK VOLUME ACCURACY – USING A PROPOFOL LABEL SET THE

                            INFUSION RATE TO 50 MCG/KG/MIN, BODY

                            WIEGHT TO 100 KG AND BOLUS TO 1000 MCG/KG.

                            INITIATE THE BOLUS AND VERIFY THAT THE

                            VOLUME MEASURED IS 10 ML +/- 3 %. 
     

    CHECK BOLUS DURATION – SET BOLUS TO 1000 MCG/KG, INITIATE A

                            BOLUS DOSE AND VERIFY THE DELIVERY

                            DURATION IS 1 MINUTE +/- 6 %.